Resistivity 0.00-20.00 MΩ • cm, 0.01 MΩ • cm resolution, +/-1% Accuracy(+/- 1 digit)
Temperature Range: -30 °C to 130 °C, 0.1 °C resolution, 0.2 °C accuracy
Calibration Mode: Automatic with built-in standard solutions manual cell constant adjustment
Cell Constants: 0.01, 0.05, 0.1, 0.5, 10/cm fixed or freely selectable 0.0080 – 19.99/cm
Analog Output 1: Isolated DC 0/4 -20mA, corresponding to conductivity or resistivity. Max load 500 Ω.
Analog Output 2: Isolated DC 0/4 -20mA, corresponding to temperature output. Max load 500 Ω.
Set-Points: Contact: 240V AC, 5A max
Activate: setpoints Hi/Low, Hi/Hi, Low/Low, ON/OFF
Enclosure: NEMA 4X IP65, with push-button control pad. HWD: 96 mm (3.77”) x 132 mm (5.20”), 1/4 DIN